TECHNICAL DATA
Measuring range 0 ÷ 100% oxygen in air
0 ÷ 60 mg/l (0 ÷ 600%) oxygen in water
Accuracy ±0.1 mg/l (±1%)
Acceptable temp. of measured solution 0 – 40 °C
Temperature compensation range 0 – 40 °C
Signal of the probe for solutions (in 20 °C)
in 100% O2 saturation 20 - 25 mV
in 0% O2 saturation max 0.3 mV
Signal drift 1%/24h
Time of response T90 below 1 minute
Internal compensation yes (thermistor)
Cathode material silver
Anode material zinc
Membrane material Teflon foil
Body and membrane cap material PVC
Electrolyte KCl 0.5 mol/l
Body diameter 12,0 ± 0,5 mm
Body length 120 ± 5 mm
Body material PVC
Membrane material Teflon
Connector RCA with M16 thread